FAQs
-
Can we transfer images from a Voluson 730 BT05 system to a BT03 system?
New images are not necessarily compatible on old equipment. New items have been added to the database, ... -
It takes forever to save 4D volumes to Sonoview on the BT05 V730. What can I do?
Under Utilities>system>Network> DICOM> Sonoview Config, change Volume Comp. to lossless. -
On version 5.0.2, why when I use the zoom function along with CFM is my CINE function not available?
This is fixed in version 5.0.3, but the workaround until then is: Utility > System Setup > select a position for the "Overview Window Position" ... -
I am trying to do batch Export of AVI files that I saved to the Sonoview, why am I not seeing the option for AVI under the drop down box?
Image Cine is the one that determines the ability to perform a batch cine, unless the cine loops/AVI's are also saved as Image Cine you will not see ... -
Why it is not possible to do an Automatic Trace Measurement on Ductus Venosus?
The problem is that the measure algorithm on VOLUSON E8 will always detect Systolic and Diastolic. According to the formula ... -
How to hide Patient name to export Images for ppts?
Not possible directly on the System but available on 4D View Software.
Most viewed questions
- please, Help me - I want to install or add anew printer to my voluson e via USB , Network or Wireless lan . Can you help me the steps - Iam Expert in Windows Utilities .
- Can we transfer images from a Voluson 730 BT05 system to a BT03 system?
- It takes forever to save 4D volumes to Sonoview on the BT05 V730. What can I do?